Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.
The moorLDI2 is a class IIa device under EC directive 9342EEC14 June 1993 Medical Device Directive.
LASER SOURCE
moorLDI2-VR

Either Helium Neon Red Gas Laser 632.8nm or Red Laser Diode:635nm to 690nm ± 5nm
Accessible power 2.5mW ± 0.5mW
Beam diameter (1/e2 points) at 50cm from scanner window 1.2mm ± 0.2mm
Beam divergence 1.4 ± 0.2milliradians
Ocular Hazard Distance 25m.
All measurements include cumulative measurement uncertainties and expected increases after manufacture.
Class 3R per IEC 60825-1:2001 and complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001.
moorLDI2-IR

Infra-Red Laser Diode: 785nm ± 5nm, 2.25mW ± 0.5mW
Beam diameter (1/e2 points) at 50cm from scanner window 1.2mm ± 0.2mm
Beam divergence is 1.4 ± 0.2milliradians
Ocular Hazard Distance 20m.
All measurements include cumulative measurement uncertainties and expected increases after manufacture.
Class 3R per IEC 60825-1:2001 and complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001.
Visible Laser Diode (target beam for infrared systems): 660nm ± 10nm, 100 to 200 µW.
Beam diameter (1/e2 points) at 50cm from scanner window 1.2mm ± 0.2mm
Beam divergence is 1.4 ± 0.2milliradians
All measurements include cumulative measurement uncertainties and expected increases after manufacture
Class 1 per IEC 60825-1:2001 and complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001.
moorLDI2-II

Either Helium Neon Red Gas Laser 632.8nm or Red Laser Diode: 635nm or 690nm ± 5nm
Accessible power 0.9mW ± 0.1mW
Beam diameter (1/e2 points) at 50cm from scanner window 1.2mm ± 0.2mm
Beam divergence 1.4 ± 0.2milliradians
Ocular Hazard Distance 10m.
All measurements include cumulative measurement uncertainties and expected increases after manufacture.
Class 2 per IEC 60825-1:2001 and complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001.
PROTECTIVE EYEWEAR REQUIREMENTS
moorLDI2-VR
The nominal ocular hazard distance is 25 metres.
Operation at single wavelength visible red: wavelength range 630nm to 700nm, OD = 1.0.
moorLDI2-IR
The nominal ocular hazard distance is 20 metres.
Operation at single wavelength infrared 785nm: wavelength range 755nm to 850nm, OD = 1.5.
ENVIRONMENT CONDITIONS
Temperature: 15°C to 30°C
Humidity: 45% to 75%
Atmospheric pressure: within the range 86,0 kPa to 106,0 kPa (645mmHg to 795mmHg).
Flammable Anaesthetics: the system must not be operated in the presence of flammable anaesthetics.
BANDWIDTH
Scan rate dependent: low frequency cut-off (3db) 20Hz, 100Hz or 250Hz.
Selectable upper cut-off frequency (0.1db) 3KHz, 15KHz or 22.5KHz. Default Bandwidth in Bold.
RANGE AND SCAN AREA
At 20cm distance, Normal area = 6.6cm x 5.5cm; Large Area =13cm x 13cm.
At 100cm distance, Normal Area = 25cm x 25cm; Large Area = 50cm x 50cm.
SCAN SPEED
Scan speed is approximately 4ms/pixel, 10ms/pixel or 50ms/pixel (at maximum resolution).
Scan duration is typically 40 seconds for a 12.5cm x 12.5cm image at 64 x 64 pixel resolution, about 6 minutes for a 50cm x 50cm image at 256 x 256 pixel resolution at 4ms/pixel and 100cm distance.
SPATIAL RESOLUTION
Up to 256 x 256 pixels (actual measurements not by interpolation): 0.2mm/pixel at 20cm, 'normal scan'; 2.0mm/pixel at 100cm, 'large scan'.
LIGHTING CONDITION
Normal, ambient room lighting.
MEASUREMENTS
FLUX Accuracy: ± 10% relative to Moor Instruments moorLDI2 standard'
Precision: ± 3% of measurement value
Range: 0-5000PU
CONC Accuracy: ± 10%
Precision: ± 5% of measurement value
Range: 0-5000AU
DC Accuracy: ± 10%
Precision: ± 3%
Range: 0-5000AU
CCD CAMERA
Auto Focus, Motorised 10X zoom. 752 X 582 Pixel Resolution.
SOFTWARE
Windows based control, processing and analysis.
COMPUTER REQUIREMENTS
Only computers complying with IEC 950 may be used.
The minimum recommended PC requirements for running the moorLDI software are:
200MHz processor; 128MB RAM; 10GB HDD, 1.44MB FDD; CD ROM; mouse; keyboard; RS232; 800 x 600 display.
Analogue Camera Model: A half length PCI card slot is required for the CCD video camera frame grabber board.
Firewire Camera Model: A half length PCI card slot or a PCMCIA slot is required for the CCD video camera Firewire interface card.
* If the PC does not have an RS232 port an RS232 to USB converter is needed, as the serial output from the moorLDI control unit is via RS232. A suitable converter can be supplied by Moor Instruments or can be purchased direct from the manufacturer at www.mct.com.tw/ (part no. U232-P9).
Operating system - Windows 2000 and XP or higher.
ELECTRICAL SAFETY CLASSIFICATION
Type of protection against electric shock - Class I.
Degree of protection against electric shock - Type B.
Degree of protection against ingress of liquid - IPXO (not protected).
Degree of protection against flammable anaesthetics - equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide.
Mode of operation - continuous.
GENERAL
Universal voltage switch mode power supply.
Range 84 to 264V AC 50VA, 50 to 60Hz.
Control Unit: Dimensions W H D mm 305 x 115 x 260: Weight 4.5kgs.
Scan Head: Dimensions W H D mm 426 x 244 x 300: Weight 8kgs.
Operating temperature: 15-30°C.
STORAGE AND TRANSPORTATION CONDITIONS
Temperature: 0-45°C.
Humidity: 0-80% RH.
Atmospheric Pressure: 500 hPa - 1060hPa.