moorVMS-HEAT
Specifications
Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.
Measured parameters: Temperature, 2 channels
Temperature measurement: Range: 5 - 50°C
Accuracy 5 - 50°C: VHP1 and VHP2 ± 0.3°C, VHP3 ± 0.5°C
Precision: ± 0.1°C
Resolution: 0.1°C
Temperature control: Range: 20°C (or ambient) to 45°C
Maximum heater power: 5 Watts
Maximum rate of heating: Limited by probe size
Medical Devices Directive classification: Class IIa (Active device for diagnosis)
GENERAL
Universal voltage switch-mode power supply range 100-230v at 30VA, 50 to 60 Hz
Dimensions W H D mm
moorVMS-HEAT 235 x 80 x 200
Weight kgs
moorVMS-HEAT 1.3
Mode of operation: continuous
Operating temperature: 10 - 30°C
STORAGE AND TRANSPORT CONDITIONS
Temperature 0 - 45°C
Humidity 0 - 80% RH
Atmospheric pressure 500-1060hPa
Type of protection against electric shock - Class 1
Degree of protection against electric shock - Type BF
Degree of protection against ingress of liquid - VHP3 probe IPX7. Instrument, all other probes IPXO (not protected)
Degree of protection against flammable anaesthetics - equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with nitrous oxide